ProScan Proscan-2000: crackling noise during operation
What happened
Crackling and abnormal noise heard while the unit was running; no error codes were shown.
What was checked
- Inspect the full wiring run from the board to each limit switch, looking for damaged insulation and spots where bare conductors can touch each other.
- Check the solder joints on the limit switch terminals; in this case the engineers found a joint with roughly a centimetre of stripped wire (as reported in the case) that could bridge to the neighbouring terminal.
Cause
In this case the cause was a defective solder joint at the limit switch: the insulation had been stripped back about a centimetre (as reported in the case), and the bare conductor was shorting to the adjacent terminal, which produced the crackling.
What was done
The engineers redid the defective joint and corrected the wiring to the microswitch; the crackling stopped after the repair. They also recommended keeping flexible MGTF-type hook-up wire on hand and, when insulation is damaged, replacing the whole board-to-microswitch run rather than patching it.
Safety
Isolate the unit from the mains and confirm absence of voltage before touching switch wiring or soldering, and complete the electrical safety test before returning the machine to clinical use.
Lesson: A vague complaint like crackling or noise can trace back to a simple solder or insulation fault at a limit switch, so check the basic wiring before suspecting larger assemblies.
Covered in the course: X-ray Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.