Early access: courses are being reviewed by field engineers
Service EngineerAcademy
Academy / Field fault notes / Puritan Bennett 760
Puritan Bennett · 760

Puritan Bennett 760: error message at start-up self-test

What happened

The ventilator displayed an error message when the start-up self-test was initiated; the exact text of the message was not recorded in the case.

What was checked

Cause

In this single case, the engineers concluded that discharged or faulty backup batteries were by far the most likely cause, with the rod position limit sensors and memory chips treated as much less likely alternatives. The conclusion is probable and was not fully confirmed in the case.

The cause is probable and was not fully confirmed.

What was done

The reported fix was to replace the batteries, after which the pre-use verification tests had to be completed — a procedure from the manufacturer's service documentation, done by authorised personnel. Without passing them, the ventilator will not start for clinical use.

Safety

A ventilator must not be returned to clinical use until the manufacturer's pre-use verification and an electrical safety test have been completed and passed, and remember that batteries store energy even with the unit switched off and disconnected from mains.

Lesson: In this case a flat backup battery presented like a complex electronics fault, so check the simple power sources first before chasing sensors and memory.

Covered in the course: Ventilator Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.

← All notes