Puritan Bennett · 840

Puritan Bennett 840: Measured volume below set value

What happened

In one reported case on a Puritan Bennett 840 ventilator, the measured exhaled volume was below the set value by about 70 units (as reported in the case), with only an occasional low exhaled volume warning. The short internal self-test passed and no error codes appeared.

What was checked

Cause

The exact cause was not determined in this case. A small calibration drift of the flow sensor or exhalation valve was considered the likely explanation, since the volume read low without triggering any error codes.

The cause is probable and was not fully confirmed.

What was done

The recommended fix was a flow sensor and exhalation valve calibration, done as a procedure from the manufacturer's service documentation by authorised personnel. As reported in the case, it requires a stable air and oxygen supply of about 5 bar and a short test hose of roughly 60–80 cm; the calibration then runs automatically and reports pass/fail. Since the calibration was a recommendation and the result was not confirmed, the outcome should be treated as probable.

Safety

A ventilator is life-support equipment: before returning it to use, run the manufacturer's full self-test and verify delivered volumes on a test lung with alarms active, and confirm electrical safety after any internal work.

Lesson: When a measured value drifts slightly with no error codes, check sensor calibration on a test lung before suspecting deeper pneumatic or circuit faults.

Covered in the course: Ventilator Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.