Puritan Bennett 840: "settings locked", ventilation will not start
What happened
After power-up the ventilator displayed a "settings locked" message and refused to start ventilation until it was put through its self-test again.
What was checked
- Ask how the unit was last switched off; as reported in the case, an uncontrolled shutdown left the 840 demanding a repeat of its power-up self-test (PVT) before it would ventilate.
- Run the PVT and record which stages fail — the engineers found errors on the air and oxygen presence sensors, the compressor and the battery.
- Inspect the battery and check the internal compressor under load, since in this case these were the components the self-test kept flagging.
Cause
Reported as probable in the case: an uncontrolled shutdown left the ventilator requiring a fresh power-up self-test, and the failing stages pointed to an aged battery and a compressor in poor condition.
The cause is probable and was not fully confirmed.
What was done
The fix was a repeat of the power-up self-test — a procedure from the manufacturer's service documentation, done by authorised personnel. The engineers treated battery replacement and compressor service as the proper way to make the test pass on real hardware; workarounds that hide a failing component from the test were discussed in the case but are not a repair.
Safety
A ventilator is life-support equipment: keep it out of patient service until it has passed the full manufacturer self-test and an electrical safety check, and isolate it from mains power before any internal work.
Lesson: When a ventilator locks its settings at start-up, it is usually demanding its self-test — treat the failed stages as real hardware faults to repair, not obstacles to bypass.
Covered in the course: Ventilator Service — Level 2: Fault-Finding →