Quertimed ETS-60: CO2 electrode error, slope low then negative
What happened
The analyzer kept reporting a CO2 electrode error. In diagnostics the electrode slope was low, and on later checks it went negative. Fresh electrodes and a full clean, including protein remover, did not help.
What was checked
- Run the electrode diagnostic and read the CO2 slope as reported in the case.
- Fit new CO2 electrodes; the slope stayed low, so the electrodes were not the root cause.
- Flush the electrode and measuring path, including protein removal; no lasting improvement.
- Trace the gas supply to the valve block and check whether calibration gas actually reaches the CO2 channel.
Cause
The engineers found a collapsed tube on the gas valves in this case: CO2 was not being supplied to the electrode at all, so the diagnostic saw a missing signal and the slope dropped low, then negative.
What was done
Restore the gas supply: clear the collapsed tube at the gas valves and confirm CO2 delivery to the electrode, then recalibrate. In this case the error cleared and the analyzer was returned to use.
Safety
Fluid lines and the measuring chamber may hold patient samples: treat them as biohazardous, wear gloves and eye protection when opening gas and fluid paths, and complete the electrical safety test before returning the analyzer to use.
Lesson: On gas analyzers, confirm that calibration gas actually reaches the electrode before condemning the electrode itself — a collapsed supply line can look exactly like a dead sensor.
Covered in the course: Lab Analyser Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.