Quertimed ETS-60: CO2 electrode error, slope low then negative

What happened

The analyzer kept reporting a CO2 electrode error. In diagnostics the electrode slope was low, and on later checks it went negative. Fresh electrodes and a full clean, including protein remover, did not help.

What was checked

Cause

The engineers found a collapsed tube on the gas valves in this case: CO2 was not being supplied to the electrode at all, so the diagnostic saw a missing signal and the slope dropped low, then negative.

What was done

Restore the gas supply: clear the collapsed tube at the gas valves and confirm CO2 delivery to the electrode, then recalibrate. In this case the error cleared and the analyzer was returned to use.

Safety

Fluid lines and the measuring chamber may hold patient samples: treat them as biohazardous, wear gloves and eye protection when opening gas and fluid paths, and complete the electrical safety test before returning the analyzer to use.

Lesson: On gas analyzers, confirm that calibration gas actually reaches the electrode before condemning the electrode itself — a collapsed supply line can look exactly like a dead sensor.