Academy / Field fault notes / Radiometer ABL 800
Radiometer · ABL 800

Radiometer ABL 800: leak detected after disinfection

What happened

Right after a disinfection run the analyzer stopped working and flagged a leak in the hydraulic system.

What was checked

Cause

The engineers found a genuine leak in the fluid path; most likely the disinfection cycle disturbed a membrane or seal, so pressure and vacuum decayed and the analyzer read it as a leak.

What was done

With fresh tubes and membranes in place, they worked through the documented leak-check procedure; the pressure/vacuum test showed where the system lost tightness, the leaking part was corrected, a repeat check held, and the analyzer returned to normal operation. Outcome reported as solved.

Safety

Treat drained sample fluid, spent tubing and membranes as potentially infectious waste, and before returning the analyzer to patient samples confirm it passes the leak check, calibration and QC.

Lesson: A leak error appearing right after disinfection usually means a disturbed membrane or seal, so renew the fluid-path wear parts and prove the repair with the documented pressure/vacuum leak test instead of just clearing the error.

Covered in the course: Lab Analyser Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.