Radiometer ABL 800: leak detected after disinfection
What happened
Right after a disinfection run the analyzer stopped working and flagged a leak in the hydraulic system.
What was checked
- Replaced the pump tubes as the first wear parts in the fluid path.
- Replaced the glucose and lactate electrode membranes.
- Ran the leak check from the manufacturer's service documentation: pressure and vacuum are built in the modules while the pO2 electrode monitors the pressure drift.
Cause
The engineers found a genuine leak in the fluid path; most likely the disinfection cycle disturbed a membrane or seal, so pressure and vacuum decayed and the analyzer read it as a leak.
What was done
With fresh tubes and membranes in place, they worked through the documented leak-check procedure; the pressure/vacuum test showed where the system lost tightness, the leaking part was corrected, a repeat check held, and the analyzer returned to normal operation. Outcome reported as solved.
Safety
Treat drained sample fluid, spent tubing and membranes as potentially infectious waste, and before returning the analyzer to patient samples confirm it passes the leak check, calibration and QC.
Lesson: A leak error appearing right after disinfection usually means a disturbed membrane or seal, so renew the fluid-path wear parts and prove the repair with the documented pressure/vacuum leak test instead of just clearing the error.
Covered in the course: Lab Analyser Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.