Radiometer ABL flex 800: autocheck will not start after electrode replacement
What happened
After new electrodes were installed and their barcodes scanned, the analyzer refused to start the autocheck procedure.
What was checked
- Reference electrode replaced as the first suspect, as reported in the case.
- Barcodes of the new electrodes scanned and old data cleared.
- Installation sequence reviewed against the manufacturer's procedure.
- The engineers found that the barcodes had been scanned only after the electrodes were already seated.
Cause
Probable cause in this case: the installation order was broken. The barcodes of the new electrodes must be scanned before the electrodes are installed; otherwise the autocheck cannot read the codes correctly and the procedure will not start.
The cause is probable and was not fully confirmed.
What was done
The engineers repeated the electrode change in the correct order — scan the barcodes of the new electrodes first, then install them — after which the autocheck could be started. This is a probable outcome in one reported case, not a general rule.
Safety
Treat the sample path and electrode compartment as biohazardous: wear gloves, avoid contact with sample and electrolyte fluids, and decontaminate per the manual before returning the analyzer to use.
Lesson: The scan-before-install step is part of the electrode replacement procedure, not paperwork — doing the steps out of order can mimic a hardware fault.
Covered in the course: Lab Analyser Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.