Radiometer ABL700/ABL800: Analyzer fails to start at power-on

What happened

As reported in the case, the analyzer would not start and produced a fault within the first minute after switching on.

What was checked

Cause

The engineers found oxidation on two silver-plated SMD fuses behind the rear panel; the oxide layer gave a poor contact, which was enough to trip a fault at startup.

What was done

The oxidized fuse contacts were cleaned, and after powering back on the analyzer started and worked normally. The case was reported as solved.

Safety

Disconnect the analyzer from mains and let the power supply discharge before working behind the rear panel, and run the electrical safety test before returning the instrument to use.

Lesson: Contact quality counts as a real fault: oxidized fuse contacts can mimic a dead power board, so check simple connections before suspecting electronics.

Covered in the course: Lab Analyser Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.