Randox Drue: falling optical density readings

What happened

Optical density dropping across cuvettes on a clinical chemistry analyzer; readings suspected to be affected by cuvette condition.

What was checked

Cause

The engineers found that the cuvette segments had yellowed and detergent had not been ordered, so the wash station was cleaning without detergent — dirty cuvettes and wash optics, not the lamp, explained the low optical density.

The cause is probable and was not fully confirmed.

What was done

Detergent was procured, and the cuvettes and wash station were cleaned by hand; the engineers recommended further verification of the wash station afterwards. The outcome was reported as probable, not fully confirmed.

Safety

Before hand-cleaning, power down and isolate the analyzer; used cuvettes and wash fluid can carry infectious material, so wear gloves and eye protection and handle waste as biohazard.

Lesson: Low optical density is not always a dying lamp — check wash chemistry supply and cuvette condition first, because a missing detergent can mimic ageing optics.