Randox Drue: falling optical density readings
What happened
Optical density dropping across cuvettes on a clinical chemistry analyzer; readings suspected to be affected by cuvette condition.
What was checked
- Checked the lamp absorbance threshold reported in the case (about 18,000, with problems reportedly starting near 14,000) to rule out lamp ageing.
- Ran a blank test per cuvette position and reviewed the optical density delta across cuvettes.
- Inspected the wash station for overflow or underfill of wash fluid.
- Visually inspected the cuvette segments and wash station — the engineers found yellowed cuvettes and no detergent in the system.
Cause
The engineers found that the cuvette segments had yellowed and detergent had not been ordered, so the wash station was cleaning without detergent — dirty cuvettes and wash optics, not the lamp, explained the low optical density.
The cause is probable and was not fully confirmed.
What was done
Detergent was procured, and the cuvettes and wash station were cleaned by hand; the engineers recommended further verification of the wash station afterwards. The outcome was reported as probable, not fully confirmed.
Safety
Before hand-cleaning, power down and isolate the analyzer; used cuvettes and wash fluid can carry infectious material, so wear gloves and eye protection and handle waste as biohazard.
Lesson: Low optical density is not always a dying lamp — check wash chemistry supply and cuvette condition first, because a missing detergent can mimic ageing optics.
Covered in the course: Lab Analyser Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.