Rayco 225: collimator light keeps burning out
What happened
The collimator positioning light kept failing: the LED burned out repeatedly and the contact pads showed scorch marks.
What was checked
- Measured the current through the LED as described in the manual — the collimator has its own adjustable current source.
- Measured the voltage in operation: reported as about 24 V, rising to the full supply voltage when the LED failed.
Cause
The engineers found that a replacement LED chip re-soldered earlier without ballast resistors had failed; by design the collimator already contains its own current stabiliser, so the failure pointed back to how the chip was wired, not to a missing ballast in the unit.
What was done
In this case the engineers re-soldered the LED chip onto a homemade substrate and added a step-down DC-DC converter from 24 V to 12 V, rated up to 1 A as reported; the light worked again, reported as solved. The cleaner option suggested in the case was to skip the extra converter and adjust the built-in current source to the minimum sufficient current — a procedure from the manufacturer's service documentation, done by authorised personnel.
Safety
Before opening the collimator, isolate and lock out the power supply, and before returning the machine to use verify the light-field alignment against the radiation field and pass an electrical safety test.
Lesson: Before adding aftermarket converters or improvised circuitry, check whether the subsystem already has an adjustable supply and set it to the minimum sufficient current per the manual.
Covered in the course: X-ray Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.