Academy / Field fault notes / Ка-660 (вероятно Rayto, не указан явно) Ca-660
Ка-660 (вероятно Rayto, не указан явно) · Ca-660

Rayto Ca-660: PT reported, INR not calculated

What happened

The coagulation analyzer reports prothrombin time (PT) results but does not calculate INR.

What was checked

Cause

The probable cause reported in the case: after a reagent lot change, the ISI value from the new lot's package insert was not re-entered. ISI differs from lot to lot, so with an outdated value the analyzer could not compute INR, while the PT measurement itself kept working.

The cause is probable and was not fully confirmed.

What was done

Enter the current ISI value for the new reagent lot from its package insert into the analyzer's reagent settings, then verify on control material that both PT and INR are calculated.

Safety

Treat patient samples, used cuvettes and reagent waste as potentially infectious material, and complete the required electrical safety test before returning the analyzer to use.

Lesson: When a calculation depends on a lot-specific reagent parameter, a lot change is the first suspect — update parameters such as ISI with every new lot before looking for a hardware fault.