Academy / Field fault notes / Remeza КМ-24.OLD10KМ (компрессор для наркозных аппаратов GE и ИВЛ Meciks)
Remeza · КМ-24.OLD10KМ (компрессор для наркозных аппаратов GE и ИВЛ Meciks)

Remeza KM-24.OLD10KM compressor: motor will not start due to blocked condensate discharge

What happened

A Remeza compressor supplying anaesthesia machines and ventilators would not start: as reported in the case, condensate stopped dumping, moisture passed into the connected apparatus, and the residual pressure prevented the compressor motor from starting.

What was checked

Cause

The engineers attributed the fault to oxidation of the bronze muffler on the condensate discharge line: the oxidised muffler stops the drain from dumping, condensate accumulates, moisture travels into the connected apparatus, and the trapped pressure is reported to block the motor start. The outcome was probable — no completed repair was reported in the case.

The cause is probable and was not fully confirmed.

What was done

No completed fix was reported in the case. The probable remedy, based on the discussion, is to remove and clean or replace the oxidised discharge muffler, confirm that the timer-controlled drain valve dumps condensate freely, drain accumulated moisture from the air path, and verify dryer performance and unloaded motor start before the compressor is returned to use.

Safety

Isolate the compressor from the mains and fully depressurise the air system before opening the condensate drain or muffler — trapped compressed air stores enough energy to eject components.

Lesson: A compressor that 'will not start' may really have a blocked condensate drain in the air-treatment chain, so trace the moisture and pressure path before replacing the motor or the compressor itself.

Covered in the course: Ventilator Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.