Renex mobile fluorography unit: camera jerks at top of travel, stand initialization will not complete

What happened

On power-up, during stand initialization, the camera jerked as it approached the top of its vertical travel, and the initialization sequence could not finish.

What was checked

Cause

The engineers found that the counterbalance spring in the lift mechanism had been removed, so the drive hit an abnormal load change near the top of travel and the camera jerked, interrupting initialization.

What was done

The spring was refitted in its original position following the manufacturer's service documentation, and after that the initialization completed normally; no other adjustments were reported in the case.

Safety

A counterbalance spring stores mechanical energy: before working on the lift mechanism, mechanically secure the camera against falling and release or control spring tension only as described in the manufacturer's documentation, then complete a full initialization and function check before returning the unit to use.

Lesson: A fault that shows up only at the end of travel usually lives in the mechanics, so before touching limits or settings, check that springs, counterweights and balances are all present and in place.