ReneX (ДРК) · статив Moviplan IC

ReneX Moviplan IC stand: overhead touchscreen panel stuck in a boot loop

What happened

On an X-ray stand, the control panel on the tube head kept cycling between a 'loading' message and a black screen and never finished booting; no log file was written to the SD card.

What was checked

Cause

The engineers found that the control panel board itself was faulty: the panel stalled early in its boot and so never wrote anything to the card.

What was done

The panel board was replaced. In this case it was only available from the supplier as a complete assembly with the touchscreen, so the whole unit was swapped; an attempt to reuse software from another identical stand was tried and did not help. After the replacement the panel booted normally and the stand was back in service.

Safety

Isolate the stand from the mains and wait for capacitors to discharge before opening the panel housing; before returning the unit to use, verify the emergency stop and exposure terminate functions.

Lesson: A boot loop with nothing written to the log card usually points at the panel hardware itself, so rule out the bus and the storage card first and then plan for a panel replacement as a complete assembly.

Covered in the course: X-ray Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.