Academy / Field fault notes / RICHARD WOLF операционная стойка с камерой logic 4K
RICHARD WOLF · операционная стойка с камерой logic 4K

Richard Wolf Logic 4K camera tower: image artifacts during coagulation

What happened

In this case, an endoscopy tower with a Logic 4K camera showed artifacts on the monitor whenever the coagulator was active inside the cavity; the image stayed clean the rest of the time.

What was checked

Cause

The engineers found the probable cause in the endovideo camera cable itself — most likely a flex fault just before the camera head — which let interference from the coagulator reach the video path. A single case, probable outcome.

The cause is probable and was not fully confirmed.

What was done

The agreed fix was replacement of the endovideo camera cable; the replacement was still being arranged at the time of the case. Until then, the workaround was to feed the coagulator from a separate grounded outlet.

Safety

Electrosurgery injects high-frequency energy into everything bonded to the tower, so before returning the system to use, verify protective bonding continuity and intact cable shields and complete the electrical safety test.

Lesson: Image artifacts that appear only while the coagulator runs usually point to a shield or conductor fault in the video cable — prove it by substitution before blaming the camera or the environment.

Covered in the course: Endoscopy Equipment Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.