Richard Wolf Logic 4K camera tower: image artifacts during coagulation
What happened
In this case, an endoscopy tower with a Logic 4K camera showed artifacts on the monitor whenever the coagulator was active inside the cavity; the image stayed clean the rest of the time.
What was checked
- The engineers first checked all electrical connections and the grounding of the tower and its cables.
- They substituted the camera head and camera control unit with known-good ones; with the other set the problem disappeared, which cleared the camera electronics.
- They flexed and twisted the camera cable just before the camera head, looking for a hidden flex break that shows up only in certain positions.
- A practical suggestion from the case: power the coagulator from a separate grounded wall outlet instead of the tower's outlets.
Cause
The engineers found the probable cause in the endovideo camera cable itself — most likely a flex fault just before the camera head — which let interference from the coagulator reach the video path. A single case, probable outcome.
The cause is probable and was not fully confirmed.
What was done
The agreed fix was replacement of the endovideo camera cable; the replacement was still being arranged at the time of the case. Until then, the workaround was to feed the coagulator from a separate grounded outlet.
Safety
Electrosurgery injects high-frequency energy into everything bonded to the tower, so before returning the system to use, verify protective bonding continuity and intact cable shields and complete the electrical safety test.
Lesson: Image artifacts that appear only while the coagulator runs usually point to a shield or conductor fault in the video cable — prove it by substitution before blaming the camera or the environment.
Covered in the course: Endoscopy Equipment Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.