Early access: courses are being reviewed by field engineers
Service EngineerAcademy
Academy / Field fault notes / ООО "РИК" (Русский Инженерный Клуб) Ланцет (хирургический СО2-лазер)
ООО "РИК" (Русский Инженерный Клуб) · Ланцет (хирургический СО2-лазер)

RIK Lancet: fault at power-on and pedal press while beam is present

What happened

The unit reported a fault immediately at power-on and again when the foot pedal was pressed, yet a beam was still coming out of the aperture — the interlock chain clearly not matching the actual emission status.

What was checked

Cause

As reported in the case, staff had pulled the 25-pin key connector (the designed interlock key) out of the rear panel, and the stereo jack plug of the foot pedal was falling apart — the interlock circuit lost its proper feedback, so the fault and the live beam appeared together.

What was done

The engineers seated the key connector back into the rear panel and replaced the damaged pedal jack; after that the fault no longer appeared at power-on or pedal press, and the case was closed as solved.

Safety

Treat the beam path as live during all checks — a CO2 beam is invisible, so keep the aperture pointed away from people, keep hands and reflective tools out of the path, and verify shutter and interlock function plus an electrical safety test before the laser returns to use.

Lesson: When an interlock fault shows up while the beam is still present, suspect the simplest external items first — key connectors, pedal plugs, cables — and never call the unit fixed until the interlock demonstrably controls emission.

Covered in the course: Fault-Finding Method for Medical Equipment →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.

← All notes