RoentgenProm Proscan 2000: software reinstall — calibrations at risk
What happened
The unit needed a software reinstall with a specific version (reported in the case), and the engineers first had to find where the machine keeps its calibration and configuration data so the settings would survive the reinstall.
What was checked
- Confirmed the exact software version required for the unit (as reported in the case) and that an installation package would be available.
- Traced where the calibrations and key settings live: dedicated calibration files plus registry branches under the manufacturer's key.
- Photographed the on-screen settings as an extra backup before touching anything.
Cause
The engineers found that the calibrations and tuned settings are not part of the installer: they sit in two separate calibration files and in the registry, so a plain reinstall without a backup would leave nothing to restore the unit's adjusted parameters from.
What was done
Before reinstalling, back up the calibration files tube.calib and rpp.calib, export the registry branches HKEY_CURRENT_USER\Software\RoentgenProm and HKEY_LOCAL_MACHINE\Software\RoentgenProm, and photograph the settings in the program. The software installation itself is a procedure from the manufacturer's service documentation, done by authorised personnel. After the reinstall, restore the calibration files and registry entries, then verify the unit before return to use.
Safety
On an X-ray unit, calibration files can govern exposure parameters — after restoring them, confirm the calibrations loaded correctly and run the unit's standard functional and QA checks before any return to clinical use.
Lesson: Before any software reinstall, back up calibrations and configuration — they usually live outside the installer, in dedicated files and the registry.
Covered in the course: X-ray Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.