Samsung H60: system power-cycles several times before starting
What happened
On power-up the button indicator lights, then drops out after 2–5 seconds as reported in the case; this repeats for 3–10 cycles, after which the system boots and runs normally.
What was checked
- Inspected the power supply under the start-up load: electrolytic capacitors and output voltages — the engineers found the output capacitors degraded
- Checked the cooling fans for free rotation and found one seized
Cause
Degraded electrolytic capacitors on the power supply output let the voltages sag under the inrush of start-up, so the supply dropped out and cycled until the charge built up; a seized cooling fan added heat and load inside the unit.
What was done
Replaced the failed electrolytic capacitors on the power supply output — six parts as reported in the case — and replaced the seized fan. After that the system powered up on the first attempt and ran normally; the case is closed as solved.
Safety
Unplug the system and verify the PSU capacitors are fully discharged before opening the supply, and complete the electrical safety check (protective earth continuity and leakage currents) before returning the unit to clinical use.
Lesson: Power-on cycling that ends in a normal boot usually points to a marginal power supply, so test its capacitors and output voltages before blaming the main board or software.