Early access · courses are being reviewed by field engineers
Service EngineerAcademy
Academy / Field fault notes / Samsung H60
Samsung · H60

Samsung H60: system power-cycles several times before starting

What happened

On power-up the button indicator lights, then drops out after 2–5 seconds as reported in the case; this repeats for 3–10 cycles, after which the system boots and runs normally.

What was checked

Cause

Degraded electrolytic capacitors on the power supply output let the voltages sag under the inrush of start-up, so the supply dropped out and cycled until the charge built up; a seized cooling fan added heat and load inside the unit.

What was done

Replaced the failed electrolytic capacitors on the power supply output — six parts as reported in the case — and replaced the seized fan. After that the system powered up on the first attempt and ran normally; the case is closed as solved.

Safety

Unplug the system and verify the PSU capacitors are fully discharged before opening the supply, and complete the electrical safety check (protective earth continuity and leakage currents) before returning the unit to clinical use.

Lesson: Power-on cycling that ends in a normal boot usually points to a marginal power supply, so test its capacitors and output voltages before blaming the main board or software.

Covered in the course: Ultrasound Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.

← All notes