Samsung HS70: poor Doppler vessel visualization with the linear probe in obese patients
What happened
On a new Samsung HS70, the LA3-16A linear probe gave weak vessel visualization on Doppler in obese patients and in the neck region; flow was sometimes only seen when an abdominal probe was used instead. No error codes were reported.
What was checked
- Checked the active preset and image settings — depth, TGC, overall gain — and confirmed the correct vascular program was in use.
- Clarified with the user whether the complaint was Doppler sensitivity or grayscale contrast.
- As reported in the case, tested penetration and Doppler resolution on the working preset with a tissue phantom.
- Turned triplex off (simultaneous 2D, color and PW) to see whether flow detection improved.
- Considered switching to lower-frequency linear probes (2–9 MHz or 3–12 MHz, as reported in the case).
Cause
The engineers found no hardware defect in the case. They attributed the weak flow signals to the limited penetration of a high-frequency linear probe in heavier patients and in deep neck vessels — an application and probe-choice limitation rather than a fault.
What was done
No repair or parts replacement was needed. The engineers recommended scanning such patients and deep neck vessels with a lower-frequency linear probe (2–9 MHz, as reported in the case) instead of the high-frequency LA3-16A; in this case the switch was reported to solve the complaint, with the improvement expected from better penetration rather than separately re-measured.
Safety
Before returning the system to use, inspect the probe lens, housing and cable for damage and perform the electrical safety checks required by the service documentation — damaged probe insulation can put the patient at risk.
Lesson: Weak Doppler in deep vessels is often a probe-choice issue, not a fault — the engineers found that confirming performance on a phantom first and then matching probe frequency to the patient resolves it fastest.
Covered in the course: Ultrasound Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.