Samsung Medison Affinity 50: touch panel drops out intermittently
What happened
On an Affinity 50 the touch control panel intermittently stopped responding; a reboot brought it back. No error codes were reported.
What was checked
- Software version checked and reported as below 5.0 in this case.
- Windows Device Manager inspected for USB power-saving settings on the touch device.
- Possible degradation of tantalum capacitors inside the panel was raised as a possibility but not confirmed.
Cause
Probable cause in this case only: on software versions below 5.0, Windows put the touch controller into USB power-saving mode and it dropped out until a reboot. This was a suspicion in one case, not a confirmed rule.
The cause is probable and was not fully confirmed.
What was done
In Windows Device Manager, on the USB entry for the touch device, the option letting the system turn the device off to save power was unchecked. The outcome is probable: with an intermittent fault, only longer observation can confirm the fix held.
Safety
Disconnect the system from the mains before any work inside the console, and run the electrical safety test before returning it to clinical use.
Lesson: Intermittent USB device dropouts with no error codes deserve a power-management settings check before any hardware is ordered.
Covered in the course: Ultrasound Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.