Samsung Medison · SonoAce X8

Samsung Medison SonoAce X8: fans run for a second, then no power-up

What happened

As reported in the case: the scanner does not start from the on/standby key on the keyboard. When mains is applied, the key LED lights and the fans spin for about a second, then everything shuts down. No error codes were shown.

What was checked

Cause

The cause was not confirmed in this case. Since the computer started reliably on the bench once serviced, colleagues judged the picture consistent with a weak computer power supply or with loss of capacitance in the motherboard electrolytics, which sag under load and drop the system out within a second of power-on — so the outcome is probable only.

The cause is probable and was not fully confirmed.

What was done

Replace the computer power supply with a known-good unit first, then check the motherboard capacitors for lost capacitance (visual inspection for bulging, capacitance/ESR measurement) and replace any that have dried out. After the repair, confirm a full start from the keyboard key and a stable run, since the diagnosis in this case was only probable.

Safety

Before opening any power supply, isolate the unit from the mains and treat its electrolytic capacitors as holding a lethal charge until discharged and verified, and complete the electrical safety test (protective earth and leakage currents) before the scanner is returned to use.

Lesson: A system that dies about a second after power-on is showing a classic ageing-electrolytic pattern, so check the bulk capacitors in the supply and on the board before chasing the switch or the controller.

Covered in the course: Ultrasound Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.