Samsung Medison X300: control panel fault shortly after power-on

What happened

On every start-up the unit ran normally for roughly the first 30 seconds (as reported in the case), then an error appeared affecting the control panel and the image. No error codes were reported.

What was checked

Cause

The engineers concluded that the control panel board itself was failing: the fault showed up at a stable time after each start and neither cleaning nor warm-up influenced it, which pointed away from contacts, dust or software.

What was done

The control panel board was replaced. The unit reportedly returned to normal operation with no repeat of the error.

Safety

Before removing the panel covers, disconnect the unit from the mains and let the internal power supplies discharge, and before the scanner goes back into use, complete the full electrical safety check, including protective earth and leakage currents.

Lesson: A fault that appears at the same point after every boot and ignores cleaning and warm-up usually means the board itself is dying — confirm it by substitution instead of chasing contacts.

Covered in the course: Ultrasound Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.