Samsung (Neurologica) · E-motion

Samsung (Neurologica) E-motion: worn signal brushes on the X-ray tube

What happened

During a wear inspection after about three years in service, the engineers found the signal brushes of the X-ray tube assembly worn out and due for replacement. There were no error codes — the wear showed up on physical inspection.

What was checked

Cause

Plain mechanical wear. The signal brushes are a consumable contact running against a moving ring, and after years of rotation they simply reach end of life. Nothing else in the circuit was reported faulty.

What was done

The engineers sourced compatible aftermarket signal brushes on their own — reported at roughly $10–22 apiece — and soldered the new brushes in place of the worn ones. Per the case, the replacement set then ran three years without further wear problems. That is one case, not a rule, so match any replacement parts against the manufacturer's service documentation.

Safety

Depower the gantry and make sure all rotating parts have fully stopped before working near the brush and ring contacts, and complete the manufacturer's electrical and radiation safety checks before returning the scanner to clinical use.

Lesson: Treat signal brushes as a scheduled wear item — inspect and replace them on time instead of waiting for contact faults to appear.

Covered in the course: CT Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.