Samsung · SonoAce R6 (Resona 6)

Samsung Resona 6: slow elastography capture and no report output

What happened

Shear-wave elastography capture ran very slowly, so the patient could not hold their breath long enough for a clean acquisition, and there was no obvious way to get the elastography results into the examination report.

What was checked

Cause

In this case the slow elastography acquisition turned out to be normal behavior for this model, and a built-in shear-wave elastography report is simply not implemented on the Resona 6 — a design limitation, not a hardware or software fault.

What was done

The elastography result is saved as a screenshot so it can be inserted into the examination report; for patients who cannot hold their breath long enough, the single-shot HQ Elasto mode is used to shorten the acquisition. Outcome reported as solved.

Safety

During prolonged elastography acquisitions, keep acoustic output as low as reasonably achievable and monitor the on-screen MI and TI indices, since slow capture extends total scan time on the patient.

Lesson: Before opening a repair case, check whether the symptom is a model limitation described in the user documentation — a missing feature or a slow mode can be by design.

Covered in the course: Ultrasound Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.