Sapphire 400: dripping from analyzer tubing

What happened

A biochemistry analyzer was dripping from its tubing; no error codes were reported.

What was checked

Cause

The engineers found the likely source was worn upper seals in the washer dispenser.

The cause is probable and was not fully confirmed.

What was done

Reported fix: the worn upper seals were removed from the washer dispenser so it seated tighter, followed by a preventive chlorine flush of the unit. The outcome was probable, not confirmed, so service the dispenser seals according to the manufacturer's service documentation.

Safety

Treat all fluid lines as potentially infectious, isolate the analyzer from power and relieve fluid pressure before opening the fluidics, and run a leak check with QC before returning the unit to use.

Lesson: A slow drip from analyzer tubing usually traces to worn seals in the wash and dispense path, so check wear parts before suspecting valves and verify the repair with a documented leak test.