Sapphire 400: Drops hanging on wash station needles when the station lifts

What happened

Every time the wash station is raised, drops form and hang on the needles. No error codes were reported in the case.

What was checked

Cause

Suspected in this case, not confirmed: residue and debris in the fluid path, most likely in the check valves, combined with reagent containers standing too high — so fluid weeps or siphons onto the needles each time the station moves up.

The cause is probable and was not fully confirmed.

What was done

Flush both systems to remove deposits, set the reagent containers so the liquid level is at or below bench height, and if dripping continues, re-seat the fittings and inspect the pump seals for air ingress. The engineers found the improvement probable only — the symptom had not been fully cleared by the end of the case.

Safety

Wear gloves and eye protection when opening or flushing the fluid paths — the acid and alkaline wash solutions are corrosive — and make sure the analyzer is dry and passes an electrical safety check before it goes back into use.

Lesson: When drips appear only during station or probe movement, check gravity-related causes first — container heights, check valves, fittings and pump seals — before blaming the probes themselves.

Covered in the course: Lab Analyser Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.