Servo-i/s: Self-test fails while the compressor is running

What happened

With the compressor running, the ventilator failed its self-test: the monitor microphone could not hear the monitor speaker, and the test aborted with an error.

What was checked

Cause

The engineers attributed the failure to excessive compressor noise: a worn overhaul kit after long service made the compressor loud enough to drown out the monitor speaker, so the microphone no longer picked it up and the acoustic check failed. The cause was treated as probable, based on the noise level and the service history.

What was done

Taking the monitor off the ventilator stand let the self-test complete without an error — this confirmed the noise as the cause and was used as a diagnostic step, not a repair. The fix recommended in the case is replacing the compressor overhaul kit per the manufacturer's schedule (a 10,000-hour interval, as reported in the case), with no changes to the test itself.

Safety

Before any work on the compressor, isolate the ventilator from mains power and let all moving parts stop, and before return to use run the full self-test and verify ventilation and alarms on a test lung.

Lesson: A self-test that fails only while ancillary equipment is running points to the environment rather than the tested unit, so check noise, vibration and power-up sequence before suspecting electronics.