Early access · courses are being reviewed by field engineers
Service EngineerAcademy
Academy / Field fault notes / Севкаврентген Томос
Севкаврентген · Томос

Sevkavrentgen Tomos: table stuck at the bottom, both limit switches active

What happened

The table descended to its lower position and would not rise. In service mode the engineers found that both the top and bottom limit switches indicated active at the same time, and the service controls did not respond.

What was checked

Cause

Probable cause: while the table rested on the bottom limit switch, the controller received contradictory limit signals, so its logic blocked the elevator drive. The engineers suspected a fault in the limit-switch circuit or its wiring rather than the motor itself; in this single case the cause was not finally confirmed.

The cause is probable and was not fully confirmed.

What was done

The fix was a procedure from the manufacturer's service documentation, done by authorised personnel: the elevator was moved off the limit-switch position under controlled service conditions, after which normal operation from the console returned, and the height potentiometer and the limit-switch circuit were rechecked. As reported in the case, the outcome was probable rather than fully confirmed.

Safety

A table moving with its limit switches defeated can crush hands or a patient — keep the travel path clear, use the slowest speed, and before return to clinical use confirm that both limit switches and all interlocks work correctly and the unit passes an electrical safety test.

Lesson: When a controller reports both limit switches active at once, that is a circuit fault to be diagnosed, not an obstacle to bypass — a mechanism may only be moved with its protection defeated through the manufacturer's documented procedure by authorised personnel.

Covered in the course: X-ray Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.

← All notes