Sevkavrentgen Tomos: table stuck at the bottom, both limit switches active
What happened
The table descended to its lower position and would not rise. In service mode the engineers found that both the top and bottom limit switches indicated active at the same time, and the service controls did not respond.
What was checked
- The engineers checked the table height potentiometer signal.
- The engineers inspected the elevator drive for a skipped or dropped gear.
- The engineers read the limit-switch inputs in service mode and found both switches active at once — a state that cannot correspond to any real table position.
Cause
Probable cause: while the table rested on the bottom limit switch, the controller received contradictory limit signals, so its logic blocked the elevator drive. The engineers suspected a fault in the limit-switch circuit or its wiring rather than the motor itself; in this single case the cause was not finally confirmed.
The cause is probable and was not fully confirmed.
What was done
The fix was a procedure from the manufacturer's service documentation, done by authorised personnel: the elevator was moved off the limit-switch position under controlled service conditions, after which normal operation from the console returned, and the height potentiometer and the limit-switch circuit were rechecked. As reported in the case, the outcome was probable rather than fully confirmed.
Safety
A table moving with its limit switches defeated can crush hands or a patient — keep the travel path clear, use the slowest speed, and before return to clinical use confirm that both limit switches and all interlocks work correctly and the unit passes an electrical safety test.
Lesson: When a controller reports both limit switches active at once, that is a circuit fault to be diagnosed, not an obstacle to bypass — a mechanism may only be moved with its protection defeated through the manufacturer's documented procedure by authorised personnel.
Covered in the course: X-ray Service — Level 2: Fault-Finding →