Early access · courses are being reviewed by field engineers
Service EngineerAcademy
Academy / Field fault notes / Siemens Acuson X300
Siemens · Acuson X300

Siemens Acuson X300: recurring undocumented error

What happened

On a Siemens Acuson X300, an img15 error appeared from time to time — reported in the case as roughly once every six months — and the code is not described in the manual. After the system was moved to a different hard disk the error seemed to disappear, but later the machine also booted normally on the original drive.

What was checked

Cause

The engineers found an intermittent fault on the TI board. The board failed in a way that surfaced as a software-level error, which is why it first looked like a disk problem, and the long gaps between recurrences made the disk swap look like a successful fix.

What was done

An attempt to repair the board failed. The engineers replaced the TI board with a second-hand unit bought at an acceptable price; per the case, the error did not return and the problem was solved.

Safety

Before opening the cabinet, isolate the system from the mains and allow internal capacitors to discharge, and before the machine goes back into clinical use, complete the electrical safety test.

Lesson: An undocumented, rare error that seems to clear after a disk swap can still be a hardware fault — intermittent board failures mimic software problems, so let time confirm the fix before closing the case.

Covered in the course: Ultrasound Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.

← All notes