Academy / Field fault notes / Siemens Acuson X300
Siemens · Acuson X300

Siemens Acuson X300: shutdown error window at end of boot

What happened

On power-up, near the end of the boot sequence and before the probe is initialised, the system displays an error window reporting a system shutdown.

What was checked

Cause

The cause was not confirmed in the case. The engineers suspected the 100 V rail of the power supply (the rail as named in the case), calling it a known weak point on this model; a supply droop during the more demanding late boot stage fits the timing of the fault.

The cause is probable and was not fully confirmed.

What was done

No confirmed repair in the case; the outcome was probable. The practical step is to measure the supply output rails under load and, if the droop is confirmed, repair or replace the power supply assembly — a procedure from the manufacturer's service documentation, done by authorised personnel.

Safety

Unplug the scanner from the mains and allow the internal capacitors to discharge before any work inside the power supply, and complete a full electrical safety test before returning the unit to clinical use.

Lesson: A shutdown error appearing late in the boot sequence, before probe initialisation, points more to a weak power supply rail than to software, so load-test the supply before touching anything else.

Covered in the course: Ultrasound Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.