Siemens Advia 60: display shows only blocks at startup
What happened
On power-up the display showed nothing but filled blocks, and the analyzer reported a smart card error together with an incorrect system clock seen in diagnostics.
What was checked
- Measure the internal backup battery — it was the first suspicion in the case, since a dead cell also loses the clock.
- Inspect internal connectors around the control board and the smart card reader: check orientation and full seating, especially after any recent service.
Cause
The smart card reader connector was plugged in reversed. The case described this as a frequent occurrence on this model: the housing does not prevent wrong orientation, the reader stops communicating, and startup fails with a garbage display plus smart card and clock errors.
What was done
The engineers re-seated the smart card connector in the correct orientation, and the analyzer started normally with the correct time. No parts were replaced.
Safety
Unplug the analyzer from the mains and let internal capacitors discharge before opening it, and complete the manufacturer's electrical safety checks, including protective earth, before the unit returns to use.
Lesson: When the display shows garbage together with several unrelated startup errors, check the seating and orientation of internal connectors first, before ordering boards or parts.
Covered in the course: Lab Analyser Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.