Siemens Artis zee: hardware failure message after IAS A hard disk replacement
What happened
After the hard disk in the IAS A subsystem was replaced, the system kept reporting a hardware failure for that subsystem, and the built-in configuration check alone did not clear it.
What was checked
- After the disk swap, run the configuration check and see whether the failure message for IAS A stays on.
- Take the configuration snapshot as described in the manufacturer's service documentation and compare the configuration before and after to spot stale entries.
Cause
The engineers found no hardware fault in this case; it pointed to stale configuration data — after the disk replacement the system still held the old record for the subsystem and kept reporting it as failed.
What was done
The engineers performed the configuration snapshot procedure from the manufacturer's service documentation. The configuration updated, and as reported in the case the hardware failure message did not come back.
Safety
Since this is a radiation-emitting system, complete the manufacturer's post-repair verification and confirm that exposure controls, interlocks and the system self-test behave normally before returning it to clinical use.
Lesson: After replacing a storage disk in a subsystem, refresh and verify the recorded system configuration — a persistent hardware failure message can be stale configuration data rather than a real fault.
Covered in the course: Angiography and C-arm Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.