Siemens Avanto: local image artifact after a dropped head coil
What happened
After the head coil was dropped, a local artifact appeared on images at the area matching the damaged spot on the coil.
What was checked
- The engineers matched the artifact position on the image to the dent in the metal shield inside the coil and inspected the deformation.
- The options raised in the case were weighed: working without the shield, straightening the plate, or cutting off the bent part; removing or trimming the shield was rejected, and some colleagues initially doubted the bent shield could cause a local artifact.
Cause
The drop bent the metal shield of the head coil, and in this case the deformed area disturbed the local RF field enough to produce a visible artifact at the same location on the image.
What was done
The engineers straightened (re-aligned) the bent shield plate; in this case the artifact disappeared and the coil returned to use. The shield was neither removed nor cut.
Safety
After any drop, inspect the coil for cracks, sharp metal edges and loose parts, keep it disconnected while working on it, and never run RF transmit with a partially connected or unloaded coil, as this can cause burns and further damage.
Lesson: A purely mechanical deformation of a coil's metal shield can be the real source of a local image artifact, so after a drop, check the coil's mechanical condition before suspecting the scanner.
Covered in the course: MRI Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.