Siemens Avanto: local image artifact after a dropped head coil

What happened

After the head coil was dropped, a local artifact appeared on images at the area matching the damaged spot on the coil.

What was checked

Cause

The drop bent the metal shield of the head coil, and in this case the deformed area disturbed the local RF field enough to produce a visible artifact at the same location on the image.

What was done

The engineers straightened (re-aligned) the bent shield plate; in this case the artifact disappeared and the coil returned to use. The shield was neither removed nor cut.

Safety

After any drop, inspect the coil for cracks, sharp metal edges and loose parts, keep it disconnected while working on it, and never run RF transmit with a partially connected or unloaded coil, as this can cause burns and further damage.

Lesson: A purely mechanical deformation of a coil's metal shield can be the real source of a local image artifact, so after a drop, check the coil's mechanical condition before suspecting the scanner.

Covered in the course: MRI Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.