Siemens · Avanto

Siemens Avanto: PALI fault on all coils except the knee coil

What happened

A periodic PALI error appeared on all coils except the knee coil. The RF characteristics test failed at step 4: with the TAS 50-ohm load connected, the amplifier reportedly jumped straight to full output power (a value of about 12200 as reported in the case), ramped it down, and was cut off on the first iteration.

What was checked

Cause

The engineers found a faulty RFSU power supply — not the RF amplifier or the TAS that were suspected first. Its degraded output pushed the amplifier into the reported full-power behaviour, which the system tripped as PALI.

What was done

The RFSU power supply was ordered and replaced promptly. The fix was done per the manufacturer's service documentation by authorised personnel; after the replacement the PALI errors stopped and the RF characteristics test passed.

Safety

Before opening the RF cabinet, isolate and lock off the supply, allow the power-supply capacitors time to discharge and verify zero voltage — and take only non-ferromagnetic tools into the magnet room.

Lesson: When a fault hits every coil except one, suspect the shared supply feeding the chain before condemning the amplifier.

Covered in the course: MRI Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.