Siemens Avanto: repeated RF coil failures and static shocks at the operator position

What happened

As reported in the case, four RF coils failed one after another, operators felt noticeable shocks when touching the RF cage, and static discharges were hitting the keyboard, mouse and console.

What was checked

Cause

The engineers found ordinary linoleum and a synthetic office chair in the operator area; with very dry winter air people charged up as they moved and released the charge in a spark at the cage, keyboard, mouse and console — and that same static was quietly killing the coils.

What was done

The floor covering and the chair were replaced with antistatic versions; as reported in the case, the problem did not return for more than three years.

Safety

Any coil that has taken repeated static hits should pass the manufacturer's electrical safety checks before it goes back into service — discharge damage is not always visible on the housing or connector.

Lesson: When the same component keeps failing for no clear reason, inspect the room itself — static from flooring, furniture and dry air can damage electronics and shock people long before any error appears.