Siemens Avanto: repeated RF coil failures and static shocks at the operator position
What happened
As reported in the case, four RF coils failed one after another, operators felt noticeable shocks when touching the RF cage, and static discharges were hitting the keyboard, mouse and console.
What was checked
- Tested all the coils from the failures — they were alive, so the cause was unlikely to be one bad coil or cable.
- Shifted attention to the operating environment instead of the scanner: floor covering, furniture, clothing and how dry the air was in the operator room and at the cage.
Cause
The engineers found ordinary linoleum and a synthetic office chair in the operator area; with very dry winter air people charged up as they moved and released the charge in a spark at the cage, keyboard, mouse and console — and that same static was quietly killing the coils.
What was done
The floor covering and the chair were replaced with antistatic versions; as reported in the case, the problem did not return for more than three years.
Safety
Any coil that has taken repeated static hits should pass the manufacturer's electrical safety checks before it goes back into service — discharge damage is not always visible on the housing or connector.
Lesson: When the same component keeps failing for no clear reason, inspect the room itself — static from flooring, furniture and dry air can damage electronics and shock people long before any error appears.