Siemens Avanto: table rolls freely by hand, drive motor spins idle

What happened

The patient table can be pushed by hand and rolls freely, while the table drive motor runs without moving the table.

What was checked

Cause

The engineers found that the most probable cause was a fracture of the motor shaft at the point where the collet grips it: the clutch no longer transmitted torque, so the table freewheeled while the motor ran.

The cause is probable and was not fully confirmed.

What was done

In this case the clutch was disassembled with ordinary open-end wrenches: the shaft was held from the opposite side of the motor and the collet nut was broken loose with a thin spanner. How the shaft was then repaired or replaced is not described in the case.

Safety

Isolate and de-energise the table drive before opening the mechanism, keep in mind that a table the drive cannot hold may roll freely — especially with a patient on it — and repeat the electrical safety tests before returning the table to patient use.

Lesson: When the drive motor runs but the table moves freely by hand, look for a mechanical break in the shaft or coupling before blaming the electronics.