Siemens Avanto: table rolls freely by hand, drive motor spins idle
What happened
The patient table can be pushed by hand and rolls freely, while the table drive motor runs without moving the table.
What was checked
- Confirmed the symptom: the drive motor runs, but the table can be moved by hand — the motor spins without load, pointing to a lost mechanical connection.
- Opened the clutch mechanism of the table drive (in this case without the dedicated service tool) and inspected the coupling, collet and motor shaft.
- Checked the clutch and the shaft while the motor was spinning idle.
Cause
The engineers found that the most probable cause was a fracture of the motor shaft at the point where the collet grips it: the clutch no longer transmitted torque, so the table freewheeled while the motor ran.
The cause is probable and was not fully confirmed.
What was done
In this case the clutch was disassembled with ordinary open-end wrenches: the shaft was held from the opposite side of the motor and the collet nut was broken loose with a thin spanner. How the shaft was then repaired or replaced is not described in the case.
Safety
Isolate and de-energise the table drive before opening the mechanism, keep in mind that a table the drive cannot hold may roll freely — especially with a patient on it — and repeat the electrical safety tests before returning the table to patient use.
Lesson: When the drive motor runs but the table moves freely by hand, look for a mechanical break in the shaft or coupling before blaming the electronics.
Covered in the course: MRI Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.