Siemens Definition AS: air calibration fails with rotation drive error
What happened
Air calibration would not complete; the log showed a Rotation motor error from the gantry rotation servo drive.
What was checked
- Checked the rotation drive cables against the checklist — no faults found.
- Verified the servo drive supply — within normal limits.
- Checked the braking resistor and found it burned out.
Cause
The braking resistor, which dissipates energy when the gantry decelerates, had burned out; the servo drive registered this as a rotation motor error and air calibration could not pass.
What was done
The burned braking resistor was replaced (a replacement resistor type was considered in the case); after the replacement the drive worked normally and air calibration passed, as reported in the case.
Safety
The gantry drive holds stored energy and the braking resistor runs very hot: isolate and lock out the drive power, let the DC bus capacitors and the resistor discharge and cool before touching anything, and repeat the electrical safety checks before the scanner returns to clinical use.
Lesson: A rotation motor error is not always the motor itself — a burned braking resistor in the servo loop can stop rotation and fail air calibration, so test the whole drive chain, not just the cables and supply.
Covered in the course: CT Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.