Siemens Definition: gantry fails to boot after reconstructor software reinstall
What happened
After a reconstructor software reinstall, the system reported configuration table mismatches and then the gantry would not boot; the display showed only the setup/configuration table. No error codes were reported in the case.
What was checked
- The engineers checked whether UMAS was actually loading: as they noted for this case, if the screen does not turn blue at startup, UMAS is not booting — here the unit sat on the configuration table.
- They compared the active configuration tables against the freshly installed software, and an attempted recovery from service level pointed to a configuration problem rather than the software itself.
Cause
The engineers concluded that the reinstall had left a damaged or non-matching configuration; with that configuration in place UMAS would not load, and without it the gantry could not start.
What was done
The fix was a procedure from the manufacturer's service documentation, done by authorised personnel: a known-good system backup (the CtScanner backup) was restored. As reported in the case, the gantry then booted and the system returned to normal operation.
Safety
Work inside the gantry involves stored electrical energy — isolate power and observe discharge times before opening anything — and after any configuration restore, run the required verification and electrical safety checks before returning the scanner to clinical use.
Lesson: When a gantry stops booting right after a software change, suspect a corrupted or mismatched configuration first and check whether the low-level boot stage comes up before reinstalling anything.
Covered in the course: CT Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.