Siemens Definition: two STRATON tube part numbers, not interchangeable
What happened
On one Definition CT a question came up about two STRATON tube part numbers — whether they are interchangeable and how to tell them apart before ordering.
What was checked
- The engineers cross-checked both part numbers against the manufacturer's spare parts documentation.
- They compared the high-voltage tube ends of the two versions externally.
Cause
The two part numbers cover different system configurations: one version is for single-tube Definition systems, the other for dual-tube ones, with different HV cable lengths. As reported in the case, the parts documentation also ties each number to a system serial-number range (one up to 67699, the other from 95000).
What was done
The tubes are not interchangeable. Externally, the dual-tube version is recognised by its small HV terminal — it plugs into an adapter rather than directly into the generator. The correct part is selected by system configuration and serial range, as reported in the case.
Safety
Before touching the tube or HV connections, power the system down and follow the manufacturer's high-voltage isolation procedure — HV terminals can hold lethal stored energy, and the tube assembly is heavy enough to require two people or lifting gear.
Lesson: Visually similar tubes can differ in configuration and cabling — always match the exact part number to the system configuration and the manufacturer's parts documentation before ordering or swapping.
Covered in the course: CT Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.