Siemens · DGM (200)

Siemens DGM (200): leakage to chassis from a pinched door wire

What happened

On a dental X-ray unit the controller power supply was carrying abnormal load. With the door cold there was no fault; the problem appeared only after the door heater had been running. The engineers found a wire that had slipped out of a dried-out clamp and was being pinched between the moving door plate and the frame.

What was checked

Cause

A wire had escaped its perished clamp and lay in the door's pinch line. Heat from the door heater softened the damaged insulation, and leakage current ran to the chassis through the grease, loading the controller power supply.

What was done

The case does not give the repair steps; the outcome is reported as solved. What mattered: the wire was freed from the pinch point, the failed clamp was replaced and the harness secured clear of the moving plate, and the leakage was confirmed gone before return to use.

Safety

Isolate the unit and prove it dead before working around the door; after any insulation repair, repeat the leakage current and protective earth test before the machine goes back into service.

Lesson: A failed clamp can quietly move wiring into a pinch point, and heat-dependent insulation faults only show when the machine is warm — inspect and test it in the state it actually works in.

Covered in the course: X-ray Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.