Early access: courses are being reviewed by field engineers
Service EngineerAcademy
Academy / Field fault notes / Siemens Emotion 16
Siemens · Emotion 16

Siemens Emotion 16: patient table stops mid-travel during scan

What happened

During scanning the patient table starts moving horizontally, then stops and the test fails with an error. No specific error code was reported in the case.

What was checked

Cause

The case shows two distinct causes behind the same symptom: contamination in the table components in one instance, and a failed D301 drive control board in the other.

What was done

Reported in the case: in one instance thorough cleaning of the table assemblies was enough; in the other the D301 board was replaced. Both were reported as solved.

Safety

Before opening the table, isolate the drive from the power supply and secure it against unexpected movement, and verify the emergency stop, limit switches and electrical safety before the table returns to clinical use.

Lesson: One symptom can come from two very different causes — clean and inspect first, but verify the control electronics before you close the case.

Covered in the course: CT Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.

← All notes