Siemens Emotion 16: patient table stops mid-travel during scan
What happened
During scanning the patient table starts moving horizontally, then stops and the test fails with an error. No specific error code was reported in the case.
What was checked
- Inspect and clean the table drive components — guides, drive, position sensors and connectors. The engineers found that in one instance cleaning alone restored normal travel.
- Run diagnostics on the D301 horizontal drive board. In another instance of the same symptom the board tested faulty and replacement was needed.
Cause
The case shows two distinct causes behind the same symptom: contamination in the table components in one instance, and a failed D301 drive control board in the other.
What was done
Reported in the case: in one instance thorough cleaning of the table assemblies was enough; in the other the D301 board was replaced. Both were reported as solved.
Safety
Before opening the table, isolate the drive from the power supply and secure it against unexpected movement, and verify the emergency stop, limit switches and electrical safety before the table returns to clinical use.
Lesson: One symptom can come from two very different causes — clean and inspect first, but verify the control electronics before you close the case.