Siemens · Emotion 16

Siemens Emotion 16: VEGA power supplies start without an enable command

What happened

Uncertainty over how the VEGA 450/650 power supplies come up in this CT: from an external start command, or simply whenever 230VAC is present at the input.

What was checked

Cause

Not a fault. The behaviour is by design: in this case the VEGA 450/650 units were found to be self-starting and have no external on/off control.

What was done

The engineers found that the supplies start as soon as mains is applied at the input, even on the bench. The trimmers for the +5VDC and +15VDC channels adjust those outputs only and have no effect on power-up. It was also noted that on the Definition the mains feeds the supplies directly, again with no external control.

Safety

Powering a supply on the bench is live mains work: use a sound earthed cord, keep terminals clear of conductive surfaces, and assume the bulk capacitors stay charged after switch-off until verified discharged; repeat the electrical safety tests before the scanner returns to use.

Lesson: Before tracing a no-power complaint through control signals, confirm whether the supply is designed to self-start on mains presence — a careful bench check settles it quickly.

Covered in the course: CT Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.