Siemens Essenza: DVD drive failure — no writing to disc from the software
What happened
The built-in DVD drive failed, so exam images could no longer be written to disc from the scanner software.
What was checked
- Physically replaced the failed DVD drive with a working unit.
- Registered the new drive in the system device configuration — a procedure from the manufacturer's service documentation, done by authorised personnel.
- Restarted the system and ran a test write to disc.
Cause
Not stated in the case; the drive unit itself had failed and needed replacement.
What was done
In this case the engineers replaced the failed drive, registered the new one in the system using the manufacturer's service procedure (done by authorised personnel), restarted the system, and writing to disc from the software worked again.
Safety
Inside an MRI room the static magnetic field is always on — keep ferromagnetic tools and spare parts clear of the magnet and follow the site's MR safety zones, and isolate power before opening the unit.
Lesson: A replaced optical drive is not always plug-and-play — plan for re-registration of the new unit in the system, not just the physical swap.