Siemens Essenza: head coil artifacts from a damaged connector
What happened
Artifacts and intermittent faults appeared during brain scans with the head coil; in this case the engineers traced the trouble to the head coil connector, which turned out to be damaged.
What was checked
- Inspection of the head coil connector and its contacts for mechanical damage and wear
- Cleaning of the coil connector contacts on a quarterly schedule, as reported in the case
- Contact-quality check, with a tune-up (TU) from the service documentation when quality was no longer acceptable
Cause
The engineers found the head coil connector physically damaged; in this case repeated disconnect and connect cycles had worn the connector and its contacts, giving unreliable contact and scan artifacts.
What was done
The head coil was left permanently connected — unnecessary disconnects were stopped — and the contacts were cleaned quarterly and checked for quality, with a tune-up when needed. With this practice the problem was resolved in this case.
Safety
Inspect and clean coil connectors only with the coil unplugged from the system and no patient in the bore, keeping ferromagnetic tools and loose metal out of the exam room per MRI safety rules.
Lesson: RF coil connectors are a wear item: avoid unnecessary disconnects and keep the contacts clean and checked, otherwise artifacts will point you to the plug, not the electronics.