Siemens · Essenza

Siemens Essenza: no helium plume from the overflow pipe before refill

What happened

Before a scheduled helium refill, no cold plume came from the cryostat overflow pipe. The helium level was reported in the case at 45%. No ice was visible in the fill port, and there was still a distinct draft from the overflow, which made the problem easy to miss.

What was checked

Cause

The engineers found an ice plug in the cryostat overflow tract: the line still passed some gas, but the plug suppressed the visible cold plume at the pipe exit.

What was done

The engineers warmed the lower, bottom section of the overflow line with a thermal probe while keeping the bypass open, so gas could pass during the work. The plug melted, the plume returned, and the helium refill went ahead as planned. This is what happened in this one case, not a universal recipe.

Safety

Do not transfer helium while the vent is suspected to be blocked, as pressure can build up in the cryostat. Cryogen work requires protection against cold burns, good room ventilation against helium accumulation, and constant monitoring of cryostat pressure; stop immediately if the pressure behaves abnormally.

Lesson: A visible cold plume from the overflow is the real proof that the vent tract is clear — a felt draft or an ice-free port does not rule out a partial ice plug before a helium refill.

Covered in the course: MRI Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.