Siemens Essenza: resonance frequency drift after shim tray work

What happened

After work on the shim tray, the magnet's resonance frequency drifted and the system would not shim back to specification.

What was checked

Cause

The engineers found that the shim plates had been reinstalled incorrectly: without their plastic spacers, stacked directly on top of one another. The changed geometry distorted the field and pulled the resonance frequency off.

What was done

The plates were re-laid in the correct order with the plastic spacers in place; then the field measurement and shim iteration were carried out per a procedure from the manufacturer's service documentation, done by authorised personnel. The frequency returned to nominal — as reported in the case.

Safety

The magnet's static field is present at all times, even with the system powered down: work at the bore only with non-ferromagnetic tools and under strict MR access control.

Lesson: Reassembly details such as spacer placement are part of the calibration — rebuild shim hardware exactly as documented before you start measuring anything.