Siemens G 60S: reboot loop right after power-on
What happened
The system drops into a restart cycle almost immediately after power-on: the Siemens logo stays on screen for a long time, then goes dark and the unit reboots again.
What was checked
- Determine at which stage the restart happens — at the logo splash or once the operating system starts loading.
- Check the workstation power supply outputs for voltage sags under load.
- Remove the internal boards one at a time to localise the fault.
Cause
Not established. In the reported case a voltage sag in the computer power supply was suspected, but since the symptom disappeared when a board was pulled, a poor board contact is just as likely — the engineers could not confirm either.
The cause is probable and was not fully confirmed.
What was done
The engineers found the reboot stopped once the boards were removed and reseated. The root cause was never confirmed, so the outcome is probable rather than a proven repair: most likely reseating cleared a bad contact. They recommended watching the system closely on subsequent power-ups.
Safety
Before opening the cabinet, isolate the system from the mains and allow time for internal discharge, and after any work inside complete the manufacturer's electrical safety test before returning the unit to clinical use.
Lesson: A boot loop that stops at the logo stage is often just a bad board contact — reseat and retest before ordering parts.