Academy / Field fault notes / Siemens G 60S
Siemens · G 60S

Siemens G 60S: reboot loop right after power-on

What happened

The system drops into a restart cycle almost immediately after power-on: the Siemens logo stays on screen for a long time, then goes dark and the unit reboots again.

What was checked

Cause

Not established. In the reported case a voltage sag in the computer power supply was suspected, but since the symptom disappeared when a board was pulled, a poor board contact is just as likely — the engineers could not confirm either.

The cause is probable and was not fully confirmed.

What was done

The engineers found the reboot stopped once the boards were removed and reseated. The root cause was never confirmed, so the outcome is probable rather than a proven repair: most likely reseating cleared a bad contact. They recommended watching the system closely on subsequent power-ups.

Safety

Before opening the cabinet, isolate the system from the mains and allow time for internal discharge, and after any work inside complete the manufacturer's electrical safety test before returning the unit to clinical use.

Lesson: A boot loop that stops at the logo stage is often just a bad board contact — reseat and retest before ordering parts.

Covered in the course: Ultrasound Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.