Siemens Harmony: moving the magnet with the field up while the cryocompressor runs
What happened
Before a site move, the question on the table was whether the magnet could be shipped still magnetized, with the OR22 cryocompressor powered the whole way so the cold head never stops — no field-down procedure first.
What was checked
- The engineers looked at the risk of the compressor tripping or refusing to run under the stray field during lifting and moving.
- They planned power continuity for the trip: a diesel generator to feed the chiller and the cryosystem for the whole route.
- They also weighed the cost of a cold-head stop mid-route — helium loss with no easy recovery on the road.
Cause
Not a fault but a logistics decision: the field was left up to avoid a field-down and ramp-up operation, and the real technical risks were compressor behaviour in the stray field and any loss of power to the cryosystem en route.
What was done
As reported in the case, the magnetized module was moved a long distance with the chiller and cryosystem running off a diesel generator the whole way, and the engineers reported the compressor ran the route without tripping. On site, three-phase mains were connected, the system was shimmed and anchored, and it went back to use with the field never taken down. This worked in this one case — it is not a general method; any transport of a magnetized magnet must follow the manufacturer's transport requirements.
Safety
A magnetized magnet must be treated as a live field at all times: clear the route and lifting zone of ferromagnetic tools, cylinders and rigging, use field-rated lifting gear, and keep bystanders outside the controlled fringe-field area for the whole move.
Lesson: When a magnet travels with the field up, the whole job stands or falls on unbroken cryosystem power and field-aware rigging — lose either one on the road and a routine move becomes a helium-loss event.
Covered in the course: MRI Service — Level 2: Fault-Finding →
Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.