What happened

The host computer of an MRI scanner running Numaris 3.5 on SunOS relied on an aging SCSI hard drive (reported in the case as 9.1 GB). The engineers needed a full disk image for backup and possible system restore, but the usual imaging tools would not work with the drive.

What was checked

Cause

No hardware fault was reported. The difficulty came from the drive itself: a SunOS SCSI disk with a non-Windows layout that mainstream imaging tools either miss or treat as an empty disk, so the problem was tool mismatch, not a failing drive.

What was done

The engineers made the image with the CloneDisk utility (32/64-bit, reported in the case) through the same SCSI adapter. After the image was written back, the drive booted normally and the system came up without problems. The whole fix was choosing a sector-level utility that could handle the disk — no intervention on the scanner itself was needed.

Safety

Before removing or connecting drives in the host computer, power the system down and isolate it from the mains and take ESD precautions, and after any system restore complete the manufacturer's verification and electrical safety checks before the scanner is returned to clinical use.

Lesson: Aging SunOS SCSI system disks are best cloned with a sector-level utility and a proper SCSI host adapter, since mainstream imaging software may fail silently or show the disk as blank — so have the right adapter and tool verified before you start.

Covered in the course: MRI Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.