Siemens Harmony/Symphony: scan start hangs at 'Adjusting (Tune)'

What happened

In this reported case, on two Siemens Harmony/Symphony MRI scanners, starting a scan sometimes did nothing: the display stayed at 'Adjusting (Tune)' and the system could sit like that for about an hour, then scan normally again. No errors were logged during these hangs, and between episodes the scanners could work fine for long stretches.

What was checked

Cause

The engineers traced the fault to the RX4 board, the reconstructor communication board. The behaviour matching the picture seen with the reconstructor optics disconnected pointed to a broken data path to the reconstructor rather than a software or database problem.

What was done

Replacement of the RX4 board. In this case the replacement resolved the hangs and the scanners returned to normal operation.

Safety

Before opening the cabinet, power down and isolate the system and follow ESD precautions — amplifier and gradient circuits can hold a stored charge — and complete the manufacturer's electrical safety checks before returning the scanner to clinical use.

Lesson: A silent hang with no logged errors can still be hardware, so compare what is common across the cases and verify the physical data path (such as the reconstructor optical link) before chasing software or database causes.

Covered in the course: MRI Service — Level 2: Fault-Finding →

Written from an anonymised case shared by service engineers. It is not a manufacturer instruction and grants no authorisation. Independent note; names are trademarks of their owners.